The MegaPulse DF-80PF DSW Surge Tester is designed to conduct the test shown in AAMI DF80, App. DD, Table US.DD.1 and Figure US.DD.1 - Damped sinusoidal waveform parameters. The resulting waveform is in accordandance with those requirements.
The DF-80PF DSW is built with 50 ohms output impedance. If another impedance is desired, please contact us.
The DF-80PF DSW is designed as a laboratory instrument and can handle a duty cycle of 1 hit every 5 minutes continuous within tolerances. Other duty cycles are optionally available:
The power output of the DF-80PF DSW is 360J. A front panel switch reverses the output polarity. The output of the Defib Surge can be measured with an oscilloscope using an appropriate 10X probe. There is no BNC oscilloscope output on the Defib Surge.
Positive and negative pulses can be delivered manually using the front panel buttons, or the pulses can be set to operate under computer control using the TestMinder Megapulse option.
MegaPulse DF-80PF DSW ships with cables, graphs of theoretical and actual waveforms, and a Calibration Certificate. One-year warranty. Options include 240V input voltage, a PLC Interface Relay Board, a computer control via Testminder MegaPulse and the duty cycle options described above.