The completely solid-state EchoTherm™ IC25 Digital, Programmable, Electronic Chilling/Heating Dry Bath is versatile, reliable and easy to use. The IC25 has simple digital controls and a two-line alphanumeric display that is backlit for easy reading. The display shows both the target and actual temperatures. Additionally, it displays the timer and all programming functions.

The chilling/heating surface is a 2.875” (7.3 cm) x 4.375” (11.1 cm) aluminum plate that is very flat for optimal contact with the sample. When unloaded, the plate will go below 0°C in under two minutes. It will heat to 37°C in even less time. Generally, the unloaded surface will decrease temperature at about 3°C/minute. Heating rates on the unloaded plate are usually about 5°C/minute. The blocks sold as accessories range in size and mass depending upon the samples used. The unit can drive a block for 1.5 ml centrifuge tubes from ambient temperature to 4°C in less than 30 minutes. When used with an accessory cover these blocks can be driven below 0°C. Temperature gradients from the center to the edge of any accessory block will be less than 0.5°C at any temperature.

OPTIONAL MODELS:
XT Model: Chilling range expanded to -20°C without sample block. Recommended for holding samples steady at low temperatures without cycling.
XR Model: dual Peltier modules for extra power, faster speeds between temperatures and expanded heating to 110°C without sample block. Strongly recommended for larger loads and for cycling temperatures up and down.

  • 5-program memory
  • Energy-efficient Peltier design
  • Digital control to 1°C from -10°C to 100°C
  • Control at ambient
  • Built-in data logger
  • Count down timer (up to 30 days) with audible alarm and user settable Auto-Off
  • RS232 I/O port for data collection
  • Electronic calibration. Factory set, but user can set to any local standard
  • Small, compact size. Uses very little bench space
  • 12-volt DC operation. External power supply provided
  • 12-month warranty
  • UL, CSA and CE compliant
  • Unattended restriction digestions or ligations at elevated temperatures then reducing temperature to 4°C automatically to reduce non-specific nuclease digestion of target DNA
  • Automatically stopping reactions by having the unit reduce sample temperature after a specific reaction time
  • Producing denatured probes or single stranded DNA molecules by heating to 90°C to 95°C for a given time and rapidly reducing temperature to 4°C to prevent renaturation
  • Automatic temperature profiles to grow bacteria at 37°C, then increasing temperature to induce events, and then lowering temperature to 4°C to prevent degradation
  • Maintaining 14°C for ligation reactions
  • Maintaining 17°C for storing oocytes
  • Storing samples at ice bucket temperatures without the mess
  • Incubating samples at 37°C in 96-well assay plates or in aluminum blocks holding 0.2, 0.5 and 1.5 ml centrifuge tubes, vials, and test tubes
  • Storing enzyme or DNA libraries at your workstation
  • Enzyme reactions and deactivations
  • Cooling blood samples prior to coagulation testing
  • Freezing and storing most any sample at your workstation

Temperature:

  • Range: -10°C to 100°C (IC25), -20C to 100C (IC25XT), -10C to 110C (IC25XR)(minimum temperature possible is 30°C below ambient without sample block)
  • Readability: 1°C
  • Accuracy: ±1°C
  • Stability: ±1°C
  • Sensor Type: Platinum RTD
  • Controller Type: PID
  • Uniformity: ±0.3°C

Timer:

  • Range: 00:01 to 30 days
  • Readability: 1 second
  • Audible Alarm: Yes, alarm sounds for 1 minute. Timer counts up until UP or DOWN key is touched

Programming:

  • Programs Stored: 5
  • Individual Program Capacity: 10 Steps
  • Program Repeat: 1–99 times
  • Memory Backup: Yes, in CMOS

Data Logger:

  • Sample times: 1 sec, 1 min or 5 minutes per data point
  • Capability: 7825 data points

Electrical:

  • Heater/Chiller: Peltier, 50 watts (IC25), 100 watts (IC25XT and IC25XR)
  • Operating Voltages: From 90 to 265 volts AC into universal power supply. 12 VDC output to chiller/heater module.
  • Power Consumption: 50 watts (IC25), 100 watts (IC25XT and IC25XR) maximum
  • Power Switch: Yes

Dimensions Chilling/Heating Module:

  • Width: 6.5” (16.5 cm)
  • Depth : 8.75” (22.23 cm)
  • Height: 3.5” (8.9 cm)

Dimensions Power Supply:

  • Length: 6.5” (16.5 cm)
  • Height: 3.2” (8.1 cm)
  • Width: 2.2” (5.6 cm)
  • Weight (power supply & chiller/heater module): 8 lbs (3.6 kg)

 

Your product has been sucessfully added to the wishlist.
Name: Torrey Pines Scientific EchoTherm IC25 Digital, Programmable, Electronic Chilling/Heating Dry Bath
Quantity: 1
Cart Total: $1,585.00
Product has been added to your shopping cart.
Name: Torrey Pines Scientific EchoTherm IC25 Digital, Programmable, Electronic Chilling/Heating Dry Bath
Quantity: 1
Cart Total: $1,585.00
Processing...

Create an Account

Already have an account? Login