Instantly measure waveforms and RMS graphs from DC to AC
Note: Use with the MEMORY HiCORDER 8847. Input cables are not supplied. Please purchase the appropriate cable for the intended application.
(Accuracy at 23 ±5 °C/73 ±9 °F, 20 to 80 % rh after 30 minutes of warm-up time and zero adjustment; accuracy/product guaranteed for 1 year)
Measurement Functions | Number of channels: 2, for voltage measurement, DC/RMS selectable |
Input Connectors | Isolated BNC connector (input impedance 1MΩ, input capacitance 30pF), Max. rated voltage to earth: 300V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
Measurement Range | 5mV to 20V/div, 12ranges, full scale: 20div, (AC voltage for possible measurement/display using the memory function: 280V rms) Low-pass filter: 5/50/500Hz, 5k/100kHz |
Measurement Resolution | 1/100 of measurement range (using 12-bit A/D conversion and when installed in 8847) |
Highest Sampling Rate | 1MS/s (simultaneous sampling across 2 channels) |
Measurement Accuracy | ±0.5% of full scale (with filter 5Hz, zero position accuracy included) |
RMS Measurement | RMS amplitude accuracy: ±1% of full scale (DC, 30Hz to 1kHz), ±3% of full scale (1kHz to 100kHz), Response time: SLOW 5s (rise time from 0 to 90% of full scale), MID 800ms (rise time from 0 to 90% of full scale), FAST 100ms (rise time from 0 to 90% of full scale), Crest factor: 2 |
Frequency Characteristics | DC to 400kHz -3dB, (with AC coupling: 7Hz to 400kHz -3dB) |
Input Coupling | AC/DC/GND |
Max Allowable Input | 400V DC (the maximum voltage that can be applied across input pins without damage) |
Dimensions, Mass | 106mm (4.17in) W × 19.8mm (0.78in) H × 196.5mm (7.74in) D, 250g (8.8oz) |
Supplied Accessories | None |