DISCONTINUED ITEM-Recommended Replacement: Vaisala Indigo 520 Transmitter
The Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMT330 is designed for demanding industrial applications where stable measurement and wide customization is important.
Vaisala HUMICAP® Performance
The HMT330 series incorporates Vaisala’s 30 years of experience in industrial humidity measurement. An updated, even more stable and chemical resistant HUMICAP® sensor is available, next to the proven reliable and accurate sensor.
Chemical purge minimizes effects of contaminants
In environments with high concentrations of chemicals and cleaning agents, chemical purge helps to maintain measurement accuracy between calibration intervals.
Chemical purge involves heating the sensor to remove harmful chemicals. The function can be initiated manually or programmed to occur at set intervals.
Graphical display of history and measurement trends
The HMT330 can be ordered with a large numerical and graphical display, which allows the user to clearly monitor operational data, measurement trends and 1-year measurement history. The optional data logger with real-time clock makes it possible to generate over four years of measured history, and zoom in on any desired time or time frame. The battery backup of the clock guarantees a reliable logging of measured data.
The display alarm allows tracking of any measured parameter, with a freely configurable low and high limit.
Data collection and (wireless) transfer to PC
The recorded measurement data can be viewed on the display or transferred to a PC with Microsoft Windows® software. The transmitter can also be connected to a network with an optional (W)LAN interface, which enables a (wireless) Ethernet connection. A USB-RJ45 cable makes it easy to connect the service port of the HMT330 to a PC.
Many ways to install and versatile outputs
Mains and DC power options, and several mounting accessories make the instrument easy to install. The HMT330 can have up to three analog outputs. Isolated galvanic power supply and analog outputs are also available. For serial interface the USB connection, RS232 and RS485 can be used. HMT330 is also capable in applying the MODBUS communication protocol and together with an appropriate connection option provides either MODBUS RTU (RS485) or MODBUS TCP/IP (Ethernet) communication.
-HMT331 for wall mounting
-HMT333 for duct mounting and tight spaces
-HMT334 for high pressures up to 100 bar and vacuum conditions
-HMT335 for high temperatures
-HMT337 for high humidity and meteorological applications
-HMT338 for pressurized pipelines up to 40 bar
Relative Humidity Measurement
Measurement Range | 0 … 100 % RH |
Accuracy at +20 °C (68 °F) | ± 1.0 % RH (0 … 90 % RH)± 1.7 % RH (90 … 100 % RH) |
Sensor | Vaisala HUMICAP® 180
Vaisala HUMICAP® 180R |
Temperature Measurement
Measurement Range | |
HMT331 | -40 … +60 °C (-40 … +140 °F) |
HMT333 | -40 … +80 °C or -40 … +120 °C
(-40 … +176 °F or -40 … +248 °F) |
HMT334/335/337/338 | -70 … +180 °C (-94 … +356 °F) |
Sensor | PT100 RTD 1/3 Class B IEC 751 |
Other variables available (depends on model) Dewpoint, temperature, mixing ratio, absolute humidity, wet bulb temperature, enthalpy, water vapor pressure.
Operating Environment
Temperature for Probe | Same as measurement ranges |
Temperature for Transmitter Body | -40 …+ 60 °C (-40 … +140 °F) |
Temperature for Transmitter Body with Display | 0 …+ 60 °C (+32 … +140 °F) |
Pressure HMT334 | 0 … 10 MPa (0 … 100 bar) |
Pressure HMT338 | 0 … 4 MPa (0 … 40 bar) |
Pressure HMT333/335/337 | vapor tight |
Inputs and Outputs
Operating Voltage with Optional Power | 10 … 35 VDC, 24 VAC |
Operating Voltage with Supply Module | 100 … 240 VAC 50/60 Hz |
Power consumption @ 20 °C (Uin24 VDC) | |
RS-232 | max. 25 mA |
Uout2 x 0 … 1 V/ 0 … 5 V/0… 10 V | max. 25 mA |
Iout2 x 0 … 20 mA V | max. 60 mA |
Analog Current Output | 0 … 20 mA, 4 … 20 mA |
Analog Voltage Output | 0 … 1 V, 0 … 5 V, 0 … 10 V |
Digital Outputs | RS-232, RS-485 (optional) |
Ethernet interface (optional) | 10/100Base-T |
Supported standards | RJ45 |
Connector | Telnet |
Protocols | |
WLAN Interface (optional) | 802.11b |
Supported Standards | RP-SMA |
Antenna Connector Type | Telnet |
Protocols | WEP 64/128,WPA |
Security | |
Authentication / Encryption | |
Open / no encryption | |
Open / WEP | |
WPA Pre shared key / TKIP | |
WPA Pre shared key / | |
CCMP (a.k.a. WPA2) | |
Optional Data Logger with Real-Time Clock | |
Logged Parameters | max. three with trend/min/max values |
Logging Interval | 10 sec (fixed) |
Max. Logging Period | 4 years 5 months |
Logged Points | 13,7 million points per parameter |
Battery Lifetime | min. 5 years |
Menu Languages | English, French, Spanish, German, Japanese, Russian, Swedish, Finnish, Chinese |
Mechanics
Probe Cable Diameter | |
HMT333 | 6.0 mm |
Other Probes | 5.5 mm |
Material (housing) | G-AlSi 10 Mg (DIN 1725) |
Housing Classification | IP65 (NEMA 4) |