Measure strain waveforms with 16-bit resolution.
Note: Use with the MEMORY HiCORDER 8847. The strain gauge converter is not provided by HIOKI, and must be purchased from a separate vendor.
(Accuracy at 23 ±5 °C/73 ±9 °F, 20 to 80 % rh after 30 minutes of warm-up time and auto-balance; accuracy/product guaranteed for 1 year)
Measurement Functions | Number of channels: 2, for distortion measurement (electronic auto-balancing, balance adjustment range within ±10000με) |
Input Connectors | Weidmuller SL 3.5/7/90G (via Conversion Cable 9769, TAJIMI PRC03-12A10-7M10.5) Max. rated voltage to earth: 33Vrms or 70V DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
Suitable Transducer
|
Strain gauge converter, Bridge impedance: 120Ω to 1kΩ, Bridge voltage: 2V ±0.05V, Gauge rate: 2.0 |
Measurement Range | 20 με to 1000 με/div, 6ranges, full scale: 20 division, Low-pass filter: 5/10/100Hz, 1kHz |
Measurement Resolution | 1/1250 of measurement range (using 16-bit A/D conversion and when installed in the 8847) |
Highest Sampling Rate | 200kS/s (simultaneous sampling across 2 channels) |
Measurement Accuracy | ±(0.5% of full scale +4με) (at 5Hz filter ON, After auto-balancing) |
Frequency Characteristics | DC to 20kHz +1/-3dB |
Dimensions, Mass | 106mm (4.17in) W × 19.8mm (0.78in) H × 196.5mm (7.74in) D, 220g (7.8oz) |
Supplied Accessories | Conversion cable 9769 × 2 (cable length 50cm/1.64ft) |