Chat with us, powered by LiveChat
Category: AccessoriesHioki
Manufacturer:

Hioki 8967 Temp Unit, 2ch, Temperature Input with Thermocouple

condition: Brand NewPart #: HI-8967Availability: Call for Availability
Request Quote

Measure temperature graphs

  • 16-bit A/D high resolution measurement
  • Isolated input, up to 300 V AC/DC from the earthNote: Use with the MEMORY HiCORDER 8847. Thermocouples are not provided by HIOKI, and must be purchased separately.

(Accuracy at 23 ±5 °C/ 73 ±9 °F, 20 to 80 % rh after 30 minutes of warm-up time and zero adjustment; accuracy/product guaranteed for 1 year)

Measurement Functions Number of channels: 2, for temperature measurement with thermocouple (voltage measurement not available)
Input Connectors Thermocouple input: plug-in connector, Recommended wire diameter: single-wire, 0.14 to 1.5mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18mm), AWG 26 to 16
Input impedance: min. 5 MΩ (with line fault detection ON/OFF), Max. rated voltage to earth: 300V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage)
Temperature Measurement Range

Note: Upper and lower limit values depend on the thermocouple

10°C/div (-100°C to 200°C), 50°C/div (-200°C to 1000°C), 100°C/div (-200°C to 2000°C), 3ranges, full scale: 20div,
Measurement resolution: 1/1000 of measurement range (using 16-bit A/D conversion and when installed in the 8847)
Thermocouple Range

(JIS C 1602-1995)

(ASTM E-988-96)

K: -200 to 1350°C, J: -200 to 1100°C, E: -200 to 800°C, T: -200 to 400°C, N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C,
B: 400 to 1800°C, W (WRe5-26): 0 to 2000°C,
Reference junction compensation: internal/ external (switchable),
Line fault detection ON/OFF possible
Data Refresh Rate 3methods, Fast: 1.2ms (digital filter OFF), Normal: 100ms (digital filter 50/60Hz), Slow: 500ms (digital filter 10Hz)
Measurement Accuracy Thermocouple K, J, E, T, N: ±0.1% of full scale ±1°C (±0.1% of full scale ±2°C at -200°C to 0°C),
Thermocouple R, S, W: ±0.1% of full scale ±3.5°C (at 0°C to 400°C or less), ±0.1% of full scale ±3°C (at 400°C or more)
Thermocouple B: ±0.1% of full scale ±3°C (at 400°C or more),
Reference junction compensation accuracy: ±1.5°C (added to measurement accuracy with internal reference junction compensation)
Dimensions, Mass 106mm (4.17in) W × 19.8mm (0.78in) H × 204.5mm (8.05in) D, 240g (8.5oz)
Supplied  Accessories Ferrite clamp × 2
Your product has been sucessfully added to the wishlist.
Name: Hioki 8967 Temp Unit, 2ch, Temperature Input with Thermocouple
Quantity: 1
Cart Total:
Product has been added to your shopping cart.
Name: Hioki 8967 Temp Unit, 2ch, Temperature Input with Thermocouple
Quantity: 1
Cart Total:
Processing...

Create an Account

Already have an account? Login