2 ch, Charge-output type piezoelectric acceleration pick-up sensor, Acceleration pick-up sensor with an internal preamp. The CHARGE UNIT 8947 is equipped with circuits for two independently usable input channels. As with previous analog units, simply slot the 8947 into the MEMORY HiCORDER for use.
Input | Number of channels: 2, Measurement objects can be selected individually for each channel. Full isolation between inputs, and between inputs and recorder. Common GND for voltage input and charge input channels. Voltage and pre-amplifier internal inputs: BNC terminals Charge input: miniature connector (#10-32 UNF) |
Suitable Converters | Charge input: piezoelectric charge output acceleration pickup sensors Internal pre-amplifier input: acceleration pickup sensors with built-in pre-amplifier |
Measurement Ranges
Charge Input (miniature connector) Pre-amplifier internal input (BNC terminal) |
50 m (m/s 2)/DIV to 10 k (m/s 2)/DIV, 12 ranges x 6 types, Measurement resolution is 1/80 to 1/32 of range (change according to measurement sensitivity), Measurement sensitivity: 0.1 pC to 10 pC/ (m/s 2), Pre-amplifier internal input measurement sensitivity: 0.1 mV to 10 mV/ (m/s 2) * When attached to recorder where full-scale (f.s.) = 20 divisions100 m (m/s 2)/DIV to 20 k (m/s 2)/DIV, 12 ranges x 6 types, 1/160 to 1/64 of range,Measurement sensitivity: 0.1 pC to 10 pC/ (m/s 2), Pre-amplifier internal input measurement sensitivity: 0.1 mV to 10 mV/ (m/s 2) * When attached to recorder where full-scale (f.s.) = 10 divisions Amplitude accuracy: ±2 % f.s. Frequency characteristics: 1 to 50 kHz +1/-3 dB |
Measurement Ranges
Voltage Input (BNC Terminal) |
500 µV to 2 V/division, 12 ranges, Measurement resolution is 1/80 to 1/32 of range (change according to measurement sensitivity) * When attached to recorder where full-scale (f.s.) = 20 divisions 1 mV to 5 V/division, 12 ranges, 1/160 to 1/64 of range * When attached to recorder where full-scale (f.s.) = 10 divisions DC amplitude accuracy: ±0.4 % f.s. Frequency characteristics: DC to 400 kHz +1/-3 dB Low-pass filter: 5/500/5 k/100 kHz Input coupling: DC, AC, GND Max. allowable input: 30 V rms or 60 V DC |
Common Functions
To all Measurement Ranges |
Anti-aliasing filter: can be turned ON or OFF, cut off frequency is the time axis, connected and set automatically by the frequency axis range |
Maximum Sampling Rate | 1 MS/s (simultaneous sampling of two channels) |
Max Rated Voltage to Earth | 30 V rms or 60 V DC (upper voltage which when applied to input channel casing or between input channels does not damage them) |
Supplied Accessories | None * The input cord is optional |